DS2422
5 of 48
DS2422 Temperature Accuracy at 3V
-2.500
-2.000
-1.500
-1.000
-0.500
0.000
0.500
1.000
1.500
2.000
2.500
3.000
3.500
-40
-30
-20
-10
0
10
20
30
40
50
60
70
80
Temperature (°C)
Error (°C)
Max. ± 0.1°C uncertainty Min. ± 0.1°C uncertainty Max. ±0.25°C uncertainty
Min. ±0.25°C uncertainty Max. ±0.5°C uncertainty Min. ±0.5°C uncertainty
Max. ±1°C uncertainty Min. ±1°C uncertainty
"Uncertainty" refers to the uncertainty of the temperature measurement when performing the 2-point calibration trim
as described in the application note. These graphs assume 11-bit temperature conversion. The accuracy can be
improved further through software correction. See the application note referenced as "Note 18" on the previous
page for details.